This article outlines some of the most common problems that can occur with your large ultrasonic cleaner. Once these potential problems have been covered, the article explains how to troubleshoot each problem before you arrange a repair.
Weak Or No Cleaning Action
Check the water level is at the correct mark, check the cleaner has had time to degas (approximately 10-15 minutes from a cold start with fresh solution). Ensure the parts are not over loaded on the basket as this can prevent even cavitation across all of the parts.
Parts Coming Out Patchy
Is part of the component not being cleaned? This is often because of air pockets. Components such as hollow washers, aerosol cans, bottle tops and cupped surfaces must be angled to allow the air to escape. Components placed on the bottom of the tank must not be touching the tank floor as sound will not travel to that area of the component. Also, make sure that there is sufficient space between articles to allow cavitation to reach all surfaces.
Cloudy Foamy Or Smelly Solution
A smelly, soapy or murky solution is a sign that the solution has gone bad, not just dirty. You can’t just top up with water and cleaning solution as the active chemicals in solution will have broken down. When replenishing solution make sure you have the correct concentration of solution and always treat cleaning chemicals and cleaning solvents with extreme care. A weak solution or a solution that is too strong can cause problems and exacerbate problems and cause irritation to skin and respiratory tract. See also Hill Sonic.
Tank Not Heating Properly

Check that the thermostat setting has not been altered. Scale build-up on the heater can cause slow heating and must be descaled regularly. Heaters also require a warm environment and can be affected by poor ventilation and cold draughts. Ensure that the unit is sited in a suitable location within the room.
Unusual Noise Or Vibration
A rattling noise can be cause for concern, but it may simply mean that a tank fixing has worked loose, often due to a cleaning bath being regularly moved from place to place. A harsh buzzing noise is typically indicative of a loose transducer mounting plate. It is also worth remembering that if the water level in the tank is below the minimum recommended for your model then it may vibrate excessively due to the lack of water cushioning the plates.
Timer Or Panel Not Responding
Make sure your power supply is turned on and the plug is pushed fully into the socket. Condensation can appear on displays that are left switched on in very damp environments and for a short while after use prevent buttons from being pressed. Try wiping the display clean with a cloth and then switching off and on again to perform a full power-down reset of the control panel.
Working through these in order will often fix the problem before it has to go for repair.